Patient Monitor Users Manual
8
of HR meter
to change in
HR
6s to 10s.
HR changes from 80bpm to 40bpm: less
than 6s to 10s.
Time to alarm
for
Tachycardia
Vent Tachycardia 1mVp-p, 206bpm:
Gain 0.5, Range 6.5 to 8.4 seconds, Average
7.2 seconds
Gain 1.0 Range 6.1 to 6.9 seconds, Average
6.5 seconds.
HR
Measurement
range
Adult: 15 bpm to 300 bpm
Pediatric and Neonatal: 15 bpm to 350 bpm
Resolution
1 bpm
Accuracy
±1% or ±1 bpm, whichever is greater
Detecting
sensitivity (II
lead)
0.20mVpp
Response time
of HR meter
to change in
HR
<12s