6 Instruction Manual TProfessional Basic 06/2011
1.3 Technical specifications
TProfessional Basic
Block
Gold anodized 96 well block or anodized 96 well XL block
Max. Heating Rate*
3.5 °C/sec
Max. Cooling Rate*
2.5 °C/sec
Avg. Heating Rate*
3.0 °C/sec
Avg. Cooling Rate*
2.0 °C/sec
Temperature Uniformity
55 °C: +/- 0.20 °C within 15 sec
70 °C: +/- 0.30 °C within 15 sec
95 °C: +/- 0.60 °C within 15 sec
Block Temperature Range
3 °C to 99 °C
Temperature Gradient **
26 °C
Gradient Temperature Range
20 to 99 °C
Control Accuracy
+/- 0.1 °C
Block Formats
96 well (70µl) or 96 well XL (100µl)
Programming
Easy spreadsheet programming
Software options
Quick start of the last five programs, view program prior to start,
toggle between spreadsheet and graphical mode, extended self
test, generate service files, adjustable ramp rates, view gradient
temperature graph, PC control over RS 232
Display
¼ VGA screen, 320 x 240 pixel
Memory
350 typical programs in up to 30 user directories (password
protected)
Auto restart after power
failure
Yes
Heated Lid
High Performance Smart Lid (HPSL)
Lid Temperature Range
30 to 99 °C
Power consumption
max. 480 Watt
Power Supply
100, 115, 230 Volt / 50 - 60 Hz
Noise emission
Very low
Interface
RS232 serial port
Dimensions (WxDxH)
28 cm x 38 cm x 24 cm
Weight
13.0 kg
Working conditions
15 °C to 35 °C, 70 % air humidity, max 2.000 NN
* Measured inside the block. ** TProfessional Basic Gradient