Manufacturer’s declaration-electromagnetic emissions
The Receiver(HR321), Receiver(HR320) is intended for use in the electromagnetic environment (for
home healthcare) specified below.
The customer or the user of the Receiver(HR321), Receiver(HR320) should assure that it is used in
such an environment.
Emission test Compliance Electromagnetic environment-guidance
(for home healthcare environment)
RF emissions
CISPR 11
Group 1 The Receiver(HR321), Receiver(HR320) uses RF
energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The Receiver(HR321), Receiver(HR320) is suitable
for use in all establishments, including domestic
establishments and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Not applicable
Manufacturer’s declaration-electromagnetic immunity
The Receiver(HR321), Receiver(HR320) is intended for use in the electromagnetic environment
(for home healthcare) specified below. The customer or the user of the Receiver(HR321),
Receiver(HR320) should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance level Electromagnetic environment-
guidance (for home healthcare
environment)
Electrostatic
discharge(ESD)
IEC 61000-4-2
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Contact:±8 kV
Air±2 kV,±4
kV,±8 kV,±15 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
+ 2kV for power
supply lines+1kV for
input/output lines
+ 2kV for power
supply lines Not
applicable
Mains power quality should be that of a
typical home healthcare environment.
Surge
IEC 61000-4-5
+ 0.5kV, +1kV line(s)
to line(s)
+ 0.5kV, +1kV,
+ 2kV line(s) to earth
+ 0.5kV, +1kV
line(s) to line(s)
Not applicable
Mains power quality should be that of a
typical home healthcare environment.