108 109
Immunity test IEC 60601
test level
Compliance level Electromagnetic environment-
guidance (for home healthcare
environment)
Voltage
Dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 25/30
cycles
Voltage interruptions:
0 % UT; 250/300
cycle
Voltage dips:
0 % UT; 0,5 cycle
0 % UT; 1 cycle
70 % UT; 30
cycles
Voltage
interruptions:
0 % UT; 300 cycle
Mains power quality should be that of a
typical home healthcare environment.
If the user of the Receiver(HR321),
Receiver(HR320) requires continued
operation during power mains
interruptions, it is recommended that the
Receiver(HR321), Receiver(HR320) be
powered from an uninterruptible power
supply or a battery.
Power
frequency(50,
60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
60 Hz
The Receiver(HR321), Receiver(HR320)
power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical home healthcare
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Manufacturer’s declaration-electromagnetic immunity
The Receiver(HR321), Receiver(HR320) is intended for use in the electromagnetic environment (for
home healthcare) specified below.
The customer or the user of the Receiver(HR321), Receiver(HR320) should assure that it is used in
such and environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-
guidance (for home healthcare
environment)
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms:
in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Portable and mobile RF
communications
equipment should be
used no closer to any part
of the Receiver(HR321),
Receiver(HR320) including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.