MEASure[:SCALar]:POWer:ALL[:DC]?
This command is used to read out the measured output power value of the 3 channels simultaneously.
Parameter type: none
Returned value: 3 channels’ actual measured output power
Example: MEAS:POW:ALL?
5.4 SCPI Command Description
5.4.1 SCPI Condition Register
You can obtain the state of the power supply and read parameters from the operation register. The different states
of the power supply can be read from 7 condition registers. These registers are status byte register, standard event
register, quest condition register and operation status register. The status byte register stores the information of 3
other registers. The following table provides the details on each register’s meaning.
bit code meaning
Operation status
sub register
0 CAL The power supply is calculating new calibration parameters.
1 UNR The status of the power supply is unregulated.
2 CV The power supply is in constant voltage condition.
3 CC The power supply is in constant current condition.
4 RI Not used
register
3 QUES If a quest enable condition changes, QUES is 1.
4
5 ESB If a standard event status enable register changes, ESB is 1.
6 MSS
7 OPER If a operation event enable register changes, OPER is 1.