CALibration:SECure[:STATe]
This command sets the protection mode to “enable” or disable. The power supply will be calibrated only if
the calibration protection is disabled and other commands are disabled, either.
Command syntax: CALibration:SECure[:STATe] <parameter 1>, <parameter 2>
Parameter 1: 0(OFF)|1(ON)
Parameter 2: calibration password
Unit: none
Reset value: 1|(ON)
*RST value: 1|(ON)
Example: CAL:SEC 0,"6322"
CALibration:SECure[:STATe]?
This command queries the protection mode: ON/OFF.
Quest syntax: CALibration:SECure[:STATe]?
Return parameter: 0(OFF)|1(ON)
Example: CAL:SEC?
CALibration:VOLTage:LEVel
This command sets the voltage calibration point.
Command syntax: CALibration:VOLTage:LEVel <parameter>
Parameter: P1|P2
Example: CAL:VOLT P1
CALibration:VOLT[:DATA]
This command reads the set voltage value of the voltage calibration point.
Command syntax: CALibration:VOLT[:DATA] <parameter>
Parameter: The actual output voltage value
Unit: V mV uV kV
Example: CAL:VOLT 1V
CALibration:CURRent:LEVel
This command sets the current calibration point.
Command syntax: CALibration:CURRent:LEVel <parameter>
Parameter: P1|P2
Example: CAL:CURR:LEV P1
CALibration:CURRent[:DATA]
This command sets the voltage value of the current calibration point.
Command syntax: CALibration:CURRent[:DATA] <parameter>
Parameter: P1|P2
Example: CAL:CURR 0.3A
CALibration:SECure:CODE
This command sets the new calibration password.