In several separate plastic containers,
prepare a small amount of:
fresh tap water
pH 7.0
pH 4.0 and/or pH 10.0 calibration
solutions.
pH 7.0 calibration
1. Rinse and place both clean pH and
Temperature probe tips in the pH 7.0
calibration solution.
2. Wait a few minutes for the probes to
come to the same temperature as
the solution.
3. Press and hold ‘CALfor three
seconds. ‘pH 7 Calibrate’ will be
displayed.
4. Press ‘CAL’ to initiate calibration.
5. Calibration is complete when all the
‘’s on the screen become solid. The
screen will display ‘OKand the pH 7’
indicator will appear to indicate
successful pH 7.0 calibration.
6. Now you can calibrate to pH 4.0
and/or pH 10.0
pH 7.0 calibration
solution
Temperature
Probe
pH Probe
water pH 7.0 pH 4.0
press cal buttonclean the pH probe
ready for pH 4 or pH 10 calibration
successful pH7 calibration