Luna
®
G3 APAP / Luna
®
G3 CPAP User Manual V1.0
47 / 48
WARNINGS!
• The device should not be used in the vicinity of other electronic equipment such
as diathermy, electrocautery and radio frequency identification (RFID), security
systems (such as electromagnetic anti-theft systems and metal detectors), cell
phone, transceiver or radio control products. If you have to do so, the device should
be observed to verify normal operation.
• Use of the device adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are
operating normally.
• Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
• Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches) to
any part of the Luna
®
G3 APAP / Luna
®
G3 CPAP, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result.
• The device may be interfered with by other equipment, even if that other
equipment complies with CISPR EMISSION requirements.
• The device may be interfered by the electromagnetic field of some known or
unknown radio frequency transmitters in the environment during use. If interference
occurs, please stay away from the interfered electromagnetic environment, or find
and turn off the electromagnetic field interference source before continuing to use
it.
• When the device is exposed to soldering, electrosurgery, defibrillation, X-ray (γ
ray), infrared radiation, transient electromagnetic field, including nuclear magnetic
resonance (MRI) and radio interference environment, the product may be
damaged.
• During operation of the device, due to electrostatic interference, the following
phenomena may occur: (1) Temporary loss of function or degradation of
performance, such as abnormal screen display, etc. The device will recover to
normal after being restarted; (2) Automatic restart of the device. These phenomena
will not affect the normal use of the device, and will not cause permanent
performance degradation or function loss of the device.