1 692 106 011 2019-03-26| Robert Bosch GmbH
Service setting – HTD 8xx | HTD 6xx / HTD 8xx | 71 en
Touch cal.
i The touch screen is calibrated here:
1. Select "TOUCH CAL." in the" SERVICE" screen .
2. The calibration is performed using three selected
points; each point must be selected three times.
3. The "SERVICE" screen appears following calibration.
Test date
i The date for the next test (calibration) is set here:
1. Select "TEST DATE" in the "SERVICE" screen.
2. Enter the date for the next test.
3. Confirm with the "OK" key.
4. Press the
key to access the "SERVICE" screen.
Preview
i The light pattern panel can be viewed and zero point
calibration performed here:
1. Select "PREVIEW" in the "SERVICE" screen.
2. The "SERVICE" screen is displayed by pressing twice
on the screen.