8. Troubleshooting list
8.1 Bonum battery operated clipping machine
8.2 Charger and Li-ion battery
Malfunction Cause Remedy
The battery-operated clipping machine does not
start
Li-ion battery not loaded Load Li-ion battery
Li-ion battery defect Insert new Li-ion battery
Li-ion battery not inserted Insert the Li-ion battery
The battery-operated clipping machine pauses dur-
ing operation
Li-ion battery deactivated through protective
circuit
Switch the battery-operated clipping machine off
and on again
Li-ion battery is empty Insert new Li-ion battery
Malfunction Detection Cause Remedy
Charger not working LED, operation indicator d does
not illuminate
Mains power cord not plugged in Plug in the mains connection line at
the mains power socket.
Li-ion battery not loaded Mains power cord defective Send charger to manufacturer for
repair; see Technical Service
Li-ion battery defect Send Li-ion battery to manufacturer
for repair, see Technical Service
Li-ion battery cannot be loaded Li-Ion battery inserted, LED,
operation indicator d does not
light up
Charger contacts dirty Clean the contacts in the charging bay,
see Reprocessing
Charger contacts damaged Send charger to manufacturer for
repair, see Technical Service
Li-ion battery defect Send Li-ion battery to manufacturer
for repair, see Technical Service
Charger defect Send charger to manufacturer for
repair, see Technical Service
Li-Ion battery inserted, LED,
warning indicator e lights up
continuously
Too high a temperature (>50°) is
measured in the Li-Ion battery dur-
ing the charging process
Remove the Li-ion battery from the
charging bay, leave to cool and repeat
the charging process
If the problem occurs again, have Li-
Ion battery inspected by manufacturer,
see Technical Service
Li-Ion battery inserted, LED,
warning indicator e blinking
Battery charging error, foreign
object on contact system or Li-Ion
battery defective
Remove the Li-Ion battery from the
charging bay, clean contact system
and repeat charging process
If the problem occurs again, have Li-
Ion battery inspected by manufacturer,
see Technical Service

Table of Contents

Other manuals for Braun Aesculap

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun Aesculap and is the answer not in the manual?

Braun Aesculap Specifications

General IconGeneral
CategoryMedical Equipment
ManufacturerBraun
MaterialStainless Steel
SterilizationAutoclavable
UsageSurgical Instruments

Related product manuals