User Manual LYNXEYE XE Detector
3-12 DOC-M88-EXX240 V1 – 03.2014
Figure 3.8: PSD real time display with PMMA sample, Cu radiation)
4. Switch to the DETECTOR plugin, set a ‘Measure time’ and start the flood field correction procedure
by pressing the ‘Measure and Set’ button.
Figure 3.9: Starting the Flood field calibration
At the end values are calculated and automatically activated and inserted in the CONFIGURATION.