User Manual LYNXEYE XE Detector
6-8 DOC-M88-EXX240 V1 – 03.2014
Note
‘Qualify Calibration’ performs a check of the actual trimming values without changes.
Figure 6.10: Result from Veto Trimming
As a result from the Veto trimming individual trim values (192) for each strip and a global DAC (Digital
Analogue Converter) Veto Threshold (here 0.028V) is determined and inserted in CONFIGURATION
and TOOLS.