User Manual LYNXEYE XE Detector
6-20 DOC-M88-EXX240 V1 – 03.2014
Figure 6.21: Create Area for Scan to determine peak position and FWHM.
The energy resolution (in eV) can be calculated by: FWHM/Obs.Max x PhotonEnergy.
In this example 0.0164/0.2108 x 8041 eV = 626 eV is achieved.
DAC Low/High Threshold should be set to 0.198V and 0.222V for full discrimination. Insert these val-
ues for this detector in the configuration for DAC Low and DAC High Threshold. The last values for the
energy discrimination are now determined: