LYNXEYE XE Detector User Manual
DOC-M88-EXX240 V1 – 03.2014 6-25
Figure 6.26: Locked coupled scan with closed discriminators showing CuKα
1/2
-, little -Kβ- and WLα
peaks
6. Load both files into EVA and evaluate the intensities for Cu-Kα1/2 (range 34.5 to 35.7 deg.)
and -Kβ peaks (range 31.2 to 32.1 deg.), save evaluation as Cu-Kbeta-suppression.eva