LYNXEYE XE Detector User Manual
DOC-M88-EXX240 V1 – 03.2014 6-27
Table 6.3: Detector properties various target materials
Tube
Energy
[eV], Kα
1,2
Mode
Gain
Energy
Resol.[eV]
Peak Pos
[V]
Kα-ratio Kβ part DAC Veto
Thr. [V]
Cr 5414 High Res. High <680 0.141-0.155
>0.75
<0.08 0.040
Co 6930 High Res. High <680 0.180-0.198
>0.75
<0.03 0.040
Cu 8048 High Res. High <680 0.209-0.241
>0.75
<0.02 0.040
Mo 17444 High Res. Low <740 0.226-0.261
>0.90
<0.01 0.026
Ag 22106 High Res. Low <780 0.286-0.332
>0.90
<0.01 0.026
Note
Width and position of the discriminators influences also the peak tails in the diffraction peak
profiles.