BSWA 308/309 BSWA Technology Co., Ltd
77
02 01 41 30 32 32 2E 38 7E 31 33 33 2E 38 2C 30 31 32 2E 38 7E 31 33 33
2E 38 2C 30 34 34 2E 38 7E 31 33 36 2E 38 03 38 0D 0A
ICPp1: Set ICCP Power
p1: ICCP power state;
0=Enable;
1=Disable;
Default: 0
Example: enable ICCP power:
02 01 43 49 43 50 30 03 29 0D 0A
Return: ACK.
ICP?: Query ICCP Power State
Example: query ICCP power state
02 01 43 49 43 50 3F 03 26 0D 0A
Return: ICCP power is enable
PR1p1_p2_p3_p4:Set Profile1
p1:
Filter;
0=A;
1=B;
2=C;
p2:
Detector;
0=Fast;
1=Slow;
2=Imp;
p3:
Integration
mode;
0=SPL;
1=PEAK;
p4:
SWN
Logger;
0=LEQ;
1=PEAK;