BÜCHI Labortechnik AG
www.buchi.com
10
New Event
To
add a new timed event, use the up/down arrows to
highlight the New button and press Enter. The New Event
window will then appear. Enter the Event Time, from 0 – 60
minutes. Enter the Event Type (Change Gain or Set Control
Relay). If Change Gain has been selected, then enter the
Gain value: 1, 2, 4, 8, or 16. If Set Control Relay has been
selected, select On or Off for the External Contact. Once the
new event has been created, press Accept to save the new
event, or press Cancel to cancel the new event.
Edit Event
To Edit an existing event, use the up/down arrows to highlight
the event to be edited from the table. Then use the right
arrow to select Edit and press Enter. The Edit Event table will
then appear, and any changes to the Event Time, Event
Type, Gain, or External Contact can be made. Press Accept
to save the event changes, or press Cancel to cancel the
changes.
Delete Event
To Delete an event, use the up/down arrows to highlight the
event to be deleted from the table. Then use the right arrow
and the down arrow to select Delete, and press Enter. The
event will then be deleted.
Once all method parameters have been entered, press
Accept to save the new method. A Save Method window will
then appear requesting a method number to be selected.
Us
e the up/down arrows or the keypad to select a method
number for the new method. Press Accept to save the new
method, or press Cancel to cancel the new method.
3.3.2 OPEN
Se
lect Method/Open and press Enter to open an existing
method. The Open Method window will then appear. Use the
up/down arrows or keypad to select a method from the list.
Press Accept and the method will then load on the main
screen, or press Cancel to cancel loading the method.
3.3.3 EDIT
To Edit an existing method, select Method/Edit and press
Enter. The Edit Method window will then appear. Use the
up/down arrows to tab through the method parameters and
enter a new Method Name, Temperature, Gas Flow, Gain,
and/or Events using the keypad. Press Accept to save the
new method settings, or press Cancel to return to the original
method settings.
3.3.4 DELETE
To Delete a method, select Method/Delete and press Enter.
The Delete Method window will then appear. Select a
method to delete from the list using the up/down arrows or by
entering/cycling a number. Press Accept to delete the
method, or press Cancel to cancel deleting the method.