3.3.5 WIZARD
The Method Wizard can help provide initial detector settings
based on the mobile phase solvent composition, mobile
phase flow rate, and sample concentration. Select
Method/Wizard and press Enter, and the first Method Wizard
screen will appear:
Select Mobile Phase Solvents
Use the left/right arrow keys to select from the list of solvents.
Use the up/down arrows to tab down the table. Enter a
percentage for each solvent. The total percentage must add
up to 100%. Once the solvents/percentages have been
entered, press Continue to continue on to the next screen in
the Method Wizard, or press Cancel to cancel using the
Method Wizard.
Mo
bile Phase Flow Rate
Select the Mobile Phase flow rate using the left/right arrows.
Possible choices are 0.2, 0.5, 1.0, 2.0, and 3.0mL/min. Press
Continue to continue on to the next Method Wizard screen, or
press Cancel to cancel the Method Wizard.
Sample Concentration
Use the left/right arrows to select the sample concentration
from the list. Press Continue to reach the next Method
Wizard screen, or press Cancel to cancel using the Method
Wizard.
Fi
nished
The final screen in the Method Wizard provides the
recommended Temperature, Gas Flow, and Gain settings
based on the mobile phase solvent composition, mobile
phase flow rate, and sample concentration. Press Finish to
continue creating the method, and the New Method window
will appear, or press Cancel to cancel the Method Wizard.
3.4 CONFIGURATION
3.4.1 ALARM
The Alarm function can be used to sound an alarm when
errors occur on the detector. Select Configuration/Alarm and
press Enter to bring up the Alarm Configuration window. Use
the left/right arrows to select Enabled or Disabled. If Enabled,
the alarm will sound when errors occur on the unit and will
continue to sound until the errors are remedied. If Disabled,
the alarm will not sound when errors occur on the unit. Press
Accept to save the alarm configuration, or press Cancel to
return to the original alarm configuration.
3.4.2 OUTPUTS
Select Configuration/Outputs and press Enter to configure the
outputs on the unit.