What to do if gas flow is lower than the set point in Buchi Security Sensors?
- DdavidbrownAug 17, 2025
If the gas flow is lower than the set point, the gas supply valve might be closed. Open the gas supply valve. The gas supply could be low or empty, so check it and replace if necessary. There may also be a blockage or leaks in the gas line before the detector; check the gas line for blockage and leaks, and remedy or replace as necessary.