Gas Flow: On/Off
Press 2 to toggle the Gas Flow On or Off. The current gas
flow rate and nebulizer gas pressure are also displayed.
Lower Drift Tube: On/Off
Press 3 to toggle the Lower Drift Tube heater On or Off. The
current Lower Drift Tube Temperature value is also displayed.
Upper Drift Tube: On/Off
Press 4 to toggle the Upper Drift Tube heater On or Off. The
current Upper Drift Tube Temperature is also displayed.
Optics Block: On/Off
Press 5 to toggle the Optics Block heater On or Off. The
current Optics Block temperature is also displayed. The
Optics Block heats to 90% of the drift tube temperature set
point.
Press Accept to return to the main screen, and the detector
will return to its original settings.
3.5.2 LOGS
VIEW ERROR LOG
Select Maintenance/Logs/View Error Log and press Enter to
access the Error Logs on the unit. Each error log entry
contains the Time, Method, Mode, and Description of the
error.
3.5.3 TESTS
Diagnostic tests can be run to help troubleshoot the unit. The
available tests are:
•
Nebulizer Gas Pressure Test
•
Gas Flow Test
•
Optics Test
•
Optics Heating Test
Select Maintenance/Tests/Nebulizer Gas Pressure Test, Gas
Flow Test, Optics Test, or Optics Heating Test and press
Enter to run a test. Refer to Section 6.2 for details on these
tests.
3.5.4 FILES
Files can be imported or exported from the unit using a pen
drive inserted into the USB Host port on the back panel of the
unit. NOTE: The pen drive must be 128MB or less; pen
drives with higher memory are NOT compatible with the 3300
ELSD.
EXPORT
To Export files from the detector, insert a pen drive into the
USB Host port on the back panel of the unit. Select
Maintenance/Files/Export/Configuration, Methods, or Error
Log and press Enter. The selected files will be exported to the
pen drive, and the pen drive containing the files can then be
removed.
IMPORT
To Import files to the detector, insert a pen drive containing
the files into the USB Host port on the back panel of the unit.
Select Maintenance/Files/Import/Configuration or Methods
and press Enter. The selected files will be imported from the
pen drive, and the pen drive can then be removed.
3.5.5 SERVICE TECH
Service Tech functions are reserved for qualified personnel
only.
3.5.6 ABOUT
Select Maintenance/About and press Enter to display the
current software versions on the unit. Press Accept to return
to the main screen.