6. DIAGNOSTICS AND TROUBLESHOOTING
6.1 ERRORS
The main screen will list any errors currently occurring on the instrument in the lower right corner. Operation error messages
will remain until the error condition is remedied. Operation errors relate to temperature and gas flow accuracy during
operation. Initialization errors will appear on bootup for 15 seconds and then disappear. Initialization errors relate to
calibration issues (i.e. if the unit is uncalibrated or if the yearly calibration is out of date).
ERRORS
error.
Drift tube temperature
reading is more than 10
°
C
(+/-) off its set point.
Heater, thermal fuse, or
electronics error.
Contact your local BUCHI
Customer Service office which
can be found on www.buchi.com.
error.
Optics block temperature
reading is more than 10
°
C
(+/-) off its set point.
Heater, thermal fuse, or
electronics error.
Contact your local BUCHI
Customer Service office which
can be found on www.buchi.com.
Gas flow reading is more than
0.3L/min (+/-) off its set point.
Gas source may be low or
empty.
Regulator pressure setting is
too low.
Nebulizer may be blocked.
Gas flow valve, meter, or
electronics error.
Check gas source and
replace if necessary.
Adjust regulator pressure to
65-80psig.
Clean nebulizer. Refer to
Section 5.2 for details.
Contact your local BUCHI
Customer Service office which
can be found on www.buchi.com.
Inlet gas pressure is low.
Inlet gas pressure is lower
than 40psig.
Regulator pressure setting is
too low.
Gas source may be low or
empty.
Gas pressure sensor or
electronics error.
Adjust regulator pressure to
65-80psig.
Replace gas source if
necessary.
Contact your local BUCHI
Customer Service office which
can be found on www.buchi.com.
and/or heaters) is out of
date.
Unit requires its yearly
recalibration.
Contact your local BUCHI
Customer Service office which
can be found on www.buchi.com.
Any errors not included in
Contact your local BUCHI
Customer Service office which
can be found on www.buchi.com.