5. If the test has failed, refer to the following table, Optics
Heating Test Results for possible causes and solutions.
OPTICS HEATING TEST RESULTS
RESUL
T
CAUSE
SOLUTION
Fail.
Detector is in
Standby mode.
Detector is still
equilibrating.
Thermal fuse may
be blown.
Optics block
heater, sensor or
other electronics
error.
Put the detector in
Run mode, allow to
equilibrate, and then
repeat the test.
Wait for the detector
to equilibrate and
then repeat the test.
Contact your local BUCHI
Customer Service office
which can be found on
www.buchi.com.
Contact your local BUCHI
Customer Service office
which can be found on
www.buchi.com.