BÜCHI Labortechnik AG Cleaning and servicing | 8
Operation Manual SyncorePlus 41/56
Use the scale marked on the tool.
u Turn the tool to adjust the unbalance compen-
sation.
Turn counterclockwise to reduce the eccentric-
ity.
Turn clockwise to increase the eccentricity.
u Remove the tool.
u Retrieve the heating plate from the pre-de-
fined stopping position. See Chapter8.3 "Mov-
ing and retrieving the heating plate into a pre-
defined position", page40
8.5 Testing the vortex movement
u Prepare the configuration. See Chapter6 "Preparing the instrument for a configu-
ration", page27
u Turn the navigation control to the rotation speed according to the requirements.
u Make sure that the requested rotation speed is the same as the rotation speed
shown on the display.