To Save a Method
Press the Scroll button until the L4 Screen appears. The screen will change to display MEMORY
LOAD, MEMORY SAVE, and DATUM OFF (or ON).
BLADE
SPEED
4250rpm
FEED
RATE
.64in/min
DISTANCE
REMAINING
1.00in
MEMORY
LOAD
MEMORY
SAVE
DATUM
OFF
L4
1. Press the MEMORY SAVE button. The screen will change to:
BLADE
SPEED
4250rpm
FEED
RATE
.64in/min
DISTANCE
REMAINING
1.00in
NEXT
METHOD
PREVIOUS
METHOD
U1
EXIT
SAVE
METHOD
The NEXT button will increase the B or U method number.
The PREV button will decrease the B or U method number.
The SAVE button will store the operating parameters at the displayed values.
The EXIT button will return to the L4 Screen.