Bunn-O-Matic Corporation
21
Step 9: Calibrate Sprayhead.
Enter Level 2 programming.
Scroll to the Calibrate Flow screen. This test is used to capture water from the sprayhead during a pre-deter
mined 60 second dispense valve cycle.
Ensure the sprayhead and funnel are in place and put a container, measuring pitcher or server, underneath the
funnel, select Yes.
Toactivatetheowratecheck,presstheBrewbutton.
The valve will open for 60 seconds. Once all of the water has dripped out, use the (-) or (+) buttons to input the
volume collected and select Done.
CALIBRATE FLOW ?
NO YES
CONTAINER RDY ?
QUIT YES
ENTER OZ: 24.5
(-) (+)DONE