Questions and Answers

  • S
    Sean DunnAug 13, 2025
    How to fix C2Dx T/PUMP TP700 Medical Equipment T/Pump® if it will not turn on?
    • R
      rperezAug 13, 2025
      If the C2Dx Medical Equipment T/Pump® Temperature Therapy Pump doesn't turn on, ensure the electrical cord is plugged into a properly grounded Hospital Grade receptacle. Visually inspect the power cord for any damage and replace it if necessary. If the cord is fine, the issue might be with the PC Board or Membrane Panel, consider replacing the Membrane Panel.
  • J
    Jacob SmithAug 17, 2025
    Why is my C2Dx T/PUMP TP700 Medical Equipment T/Pump® Temperature Therapy Pump not heating?
    • J
      jessicacurryAug 17, 2025
      If your C2Dx Medical Equipment T/Pump® Temperature Therapy Pump isn't heating, first refill the reservoir with room temperature water. Ensure there are no flow restrictions. Verify the Therapy Cycle is ON and the Set Point isn't too low. Check the power to the heater at J2 on the PCB; if power is present and the thermostat is closed, the heater is likely defective and may need replacement. If power is present but the thermostat is open, the thermostat is likely defective and should be replaced if it fails the Backup Limit Thermostat Test. If there is no power to the heater, the PC Board may need replacing.
  • J
    James WaltersAug 20, 2025
    What to do if C2Dx T/PUMP TP700 T/Pump® Temperature Therapy Pump will not pump?
    • J
      Janet MoralesAug 21, 2025
      If the C2Dx Medical Equipment T/Pump® Temperature Therapy Pump isn't pumping, first refill the reservoir with room temperature water to the proper level. Check the power across the motor connections (TP700 / TP700C - 120V ± 10%). If power is present, there may be a defective motor or impeller magnet assembly. If there is no power, replace the PC Board. Replace the motor if the fan isn't spinning when the unit is plugged in and in run mode. Otherwise, check the impeller for obstructions or replace the impeller magnet assembly. If the fan is spinning and the impeller magnet assembly is fine, check the magnet driver assembly to ensure the screw holding the driver magnet to the motor shaft is tight.
  • L
    Lisa GarciaAug 25, 2025
    Why is the flow indicator light on my C2Dx T/PUMP TP700 Medical Equipment T/Pump®?
    • S
      Summer HebertAug 25, 2025
      If the flow indicator light is on for your C2Dx Medical Equipment T/Pump®, it means water flow to the pad or hose is restricted. Check the following: - Ensure the hose is straight. - Verify the clamp is open. - Refill the reservoir with room temperature water to the proper level.
  • A
    Alisha BrownAug 28, 2025
    What does the warning indicator and audible alarm mean on my C2Dx Medical Equipment T/Pump®?
    • K
      Kimberly MillerAug 28, 2025
      The warning indicator and audible alarm on your C2Dx Medical Equipment T/Pump® can indicate several things: a high heat or cooling setpoint was selected (outside body temperature range), power loss during therapy mode (in which case the alarm will beep for about 10 minutes), the unit is timing an 'ON' therapy cycle, or the unit just entered or exited lock mode. For a power failure, insert the plug fully into the receptacle, place the unit into Standby mode, then unplug the T/Pump® Temperature Therapy Pump.The 'ON' Therapy Cycle and Lock mode indications are normal.
  • T
    Tyler FloresSep 1, 2025
    Why is the warning indicator on in Standby Mode on my C2Dx Medical Equipment unit?
    • J
      Jessica NicholsonSep 1, 2025
      If the warning indicator is on while your C2Dx Medical Equipment unit is in Standby Mode, it likely shut down due to an over-temperature condition. To resolve this, empty the reservoir and refill it with room temperature water. Ensure all clamps are open and press the On/Standby button. Verify flow through the pad; the warning light should turn off within 5 minutes.
  • K
    kimberlypayneSep 4, 2025
    What to do if temperature or therapy time buttons do not work on C2Dx T/PUMP TP700 Medical Equipment?
    • J
      Joshua BonillaSep 4, 2025
      If the Temperature or Therapy Time buttons on your C2Dx Medical Equipment aren't working, they may be locked. Press and hold the lock button for two seconds to unlock them. If that doesn't work, the membrane panel may be defective and require replacement.
  • J
    James AyalaSep 8, 2025
    How to fix water leaks from hose connectors on C2Dx T/PUMP TP700 Medical Equipment?
    • W
      William WoodardSep 8, 2025
      If water leaks from the hose connectors of your C2Dx Medical Equipment, replace the Clik-Tite® connector. Use P/N: 03887001 for the male connector and P/N: 03884001 for the female connector. Ensure the Clik-Tite® connector is shut and the pad connection to the pump is secure. If leaking continues, replace the connectors or pad if defective.

Summary

Summary of Safety Precautions

Specifications

Cleaning

Quick Reference Replacement Parts

Preventive Maintenance

Functional Checks

Procedure for performing functional tests to ensure optimal performance and safety.

Functional Check Setup - Clik-Tite Connectors

Functional Check Setup - Colder Connection

Internal Pump Inspection

Steps for inspecting the internal components of the pump for maintenance and troubleshooting.

Flow and Operating Temperature Test

Procedure to test the system's flow rate and operating temperature accuracy.

Backup Limit Thermostat Test

Water Leak Check

Current Leakage Check

Troubleshooting

Wiring Diagram

Service

Theory of Operation

Reservoir

Upper Front and Rear Housing

Procedure for removing and reinstalling the upper and rear housing components.

PC Board

Detailed steps for removing and replacing the main printed circuit board.

Heater Kit Assembly

T;Pump Temperature Therapy Pump Assembly

C2Dx T/PUMP TP700 Specifications

General IconGeneral
BrandC2Dx
ModelT/PUMP TP700
CategoryMedical Equipment
LanguageEnglish