Functional checks (Continued)
Tools required for all functional test procedures:
• TPT9 - to measure the temperature and the ow of the water entering the pad (Temperature test tool accuracy ±
• 0.2°F)
• To use Colder style quick disconnects, also order adapter hose P/N 77926000
• Stop watch
• Mul-T-Pad
®
Temperature Therapy Pad
• Test probe, P/N 100925000
• Current leakage meter
• Ground resistance meter
• Ohm meter
• #2 Phillips screwdriver
• Vacuum cleaner or air compressor
• Functional check and safety inspection form on page 15
Note: To order any of the parts listed, contact your dealer or contact C2Dx
®
customer service.
Functional check and safety inspection form
Inspection forms vary from hospital to hospital. The following sample form is intended as a guide to record the important
parameters.
T/Pump
®
Temperature Therapy Pump functional check and safety inspection form
Serial number: Date:
Model number: Check one of the boxes
Action takenItem Value
Okay,
no action
Action
needed
1
Inspect physical condition (line cord,
plug, housing)
Inspect hose connections and
connectors
2
Measure ow, >7 gph (26.5 lph) indicate
value
3
Measure operating temperature at
107°F± 2°F (42°C± 1°C) indicate average
4
Backup limit test 110°F - 120°F
(43.3° C - 49°C)
5 Water leak check
6
Measure good resistance, <0.5 ohm
indicate value
7
Measure current leakage,
<100μA (120 V)
101176 Rev 1
Preventive maintenance
15 c2dx.com