3
Summary of safety precautions
Carefully read and strictly follow the warnings and cautions listed on this page. Service only by qualied personnel.
WARNING
Always unplug the product before cleaning or disinfecting to avoid the risk of shock.
Risk of electric shock. Disconnect power before servicing or cleaning the T/Pump
®
Temperature Therapy Pump.
Only qualied service personnel should perform this Functional Check. Test procedures completed improperly may result
in equipment damage.
Do not perform functional checks with an empty reservoir as this may result in damage to the product.
Always perform a functional check after any eld service to make sure the product operates as intended.
CAUTION
Do not modify the product or any components of the product. Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids
its warranty.
Do not use quaternary disinfectant solutions on the external surface of this product.
Do not spray cleaning or disinfectant solutions onto the pump. Frequent or prolonged disinfectant solutions may result in
damage to the product.
Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
Do not place unprotected circuit boards on the oor.
Do not use the product next to or stacked with other equipment. If next to or stacked is necessary, observe the product to
verify normal operation in the conguration in which it used.
101176 Rev 1 c2dx.com