Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all C2Dx
®
products. You may need to perform preventive maintenance
checks more often based on your level of product usage. Service only by qualied personnel.
WARNING
Risk of electric shock. Disconnect power before servicing or cleaning the T/Pump
®
Temperature Therapy Pump.
Note: Change the water monthly or more often depending on use.
Inspect the following items:
Line cord, the plug, and the housing for cuts or cracked insulation and quality of the strain relief, if damaged
replace
Membrane panel for cuts or cracks, if damaged replace
Plastic parts on the pump for cracks
Connectors for cracks and missing O-rings, replace connectors if necessary
Hose connections and connectors function properly
Pads and hoses are free from tears, cracks, and water leaks
Perform Functional checks on page 14.
Inspection Value Okay Action Needed Action Taken
Measure current
leakage <100 μA
Product serial number:
Completed by:
Date:
Functional checks
Perform Functional Checks during maintenance, or after you repair an internal component, or as specied per
your facilities preventive maintenance program. This is to make sure you have optimum performance, dependability, and
safety of your product.
WARNING
Only qualied service personnel should perform this Functional Check. Test procedures completed improperly may result
in equipment damage.
Do not perform functional checks with an empty reservoir as this may result in damage to the product.
Always perform a functional check after any eld service to make sure the product operates as intended.
This section provides a complete check of all T/Pump
®
Temperature Therapy Pump functions. See (Functional check and safety
inspection form on page 15) to record test data. Follow the steps in this section. Pay attention to each step, and the expected result. If at
any time you cannot get the expected result, press the On/Standby button to stop the test, then restart the procedure. If after a second
attempt you cannot get the expected result, press On/Standby button to stop the test. Unplug the T/Pump
®
Temperature Therapy Pump
and call your dealer or contact C2Dx
®
Technical Support Department for help.
101176 Rev 1