Functional checks (Continued)
Current leakage check
Check the current leakage at this point with the unit full and connected to a pad.
Procedure:
1. Connect the pump to the Mul-T-Pad
®
Temperature Therapy Pad.
2. Straighten the pad and hoses.
3. Open the hose clamps.
4. Open the ll cap on top of the pump.
5. Fill with room temperature water to the Heating water line. Close the ll cap on the top of the pump.
6. Plug the pump into a properly grounded Hospital Grade receptacle.
Notes: The pump performs a self test of the lights and audible alarm, with a light pattern and a short audible beep.
The pump goes to Standby mode with only the Standby light on.
7. Press the On/Standby button.
8. Using the Current leakage meter, measure all combinations of heater in the on and off states by pressing the
On/Standby button.
Notes
Access to ground for current leakage testing is available through the hole where (See Figure 4 Access to ground
on page 20) the hose connects to the pump.
The highest reading should be less than 100μA.
Figure 4: Access to ground
9. Record the highest reading.
10. Disconnect the current leakage meter.
101176 Rev 1