IFU-2800 Rev E 11/18
12
Table 2 — Guidance and manufacturer’s declaration – electromagnetic immunity
The CapsoAccess® system is intended for use in the electromagnetic environment specied below.
The customer or the user of the CapsoAccess® system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast transient/
burst IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply lines
± 1 kV for input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Voltage dips:
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0% UT; 1 cycle and 70% UT; 25/30 cycles
Single phase: at 0°
Voltage interruptions:
0% UT; 250/300 cycles
Voltage dips:
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270°
and 315°
0% UT; 1 cycle and 70% UT; 25/30 cycles
Single phase: at 0°
Voltage interruptions:
0% UT; 250/300 cycles
Mains power quality should be that of a
typical commercial or hospital
environment. lf the user of the
CapsoAccess® system requires continued
operation during power mains
interruptions, it is recommended that the
CapsoAccess® system be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic elds should be
at levels characteristic of a typical location
in a typical domestic, commercial or
hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.