IFU-2800 Rev E 11/18
13
Table 3 — Guidance and manufacturer’s declaration – electromagnetic immunity
The CapsoAccess® system is intended for use in the electromagnetic environment specied below.
The customer or the user of the CapsoAccess® system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
IEC 61000-4-6
Conducted RF
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 MHz
6 Vrms in ISM bands
between 150 KHz and
80MHz
3 V/m
80 MHz to 2.7 GHz
3 Vrms
6 Vrms in ISM bands
3 V/m
Portable and mobile RF communications equipment should be used no closer
to any part of the EQUIPMENT, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency of
the transmitter. Recommended distance
d = 1.2√P 150 KHz to 80 MHz
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.7 GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from xed RF transmitters, as determined by an electromagnetic
site survey
a
, should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from structures, objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an
electromagnetic site survey should be considered. If the measured eld strength in the location in which the CapsoAccess® system is used exceeds the applicable RF
compliance level above, the CapsoAccess® system should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as re-orienting or relocating the CapsoAccess® system.
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.