26
TROUBLESHOOTING, Cont.
Error Codes, Cont.
CONFIG RQD
E
2
PROM checksum failure. Indicates improper stored calibration data, calibration is necessary.
CORRECTIVE ACTION: Recalibrate with calibrated test weight.
-ERROR-
An invalid keypad entry was attempted:
A.
PRINT
key pressed with a negative weight.
B.
TARE
key pressed to enter a push button tare value of a negative value.
C. ENTER key pressed to enter a tare weight value that exceeds the scale capacity.
D. ENTER key pressed to enter a tare weight value that is inconsistent with the scale
division value (i.e. attempt to enter a tare of 123 with scale divisions of 5).
E.
ZERO
key pressed when the gross weight is outside the scale zero weight range.
F.
lb/kg
key pressed to change to kilograms when the kilogram tare weight value exceeds
4 digits in length.
CORRECTIVE ACTION: Determine which of the reasons for the error display is applicable and
take the appropriate corrective action.
ERROR A-D
The Analog to Digital converter chip has stopped producing weight.
CORRECTIVE ACTION: Consult your scale service provider.
ERROR DRDY
The Analog to Digital converter chip is not responding.
CORRECTIVE ACTION: Consult your scale service provider.
FILE FULL
The operator is attempting to add an ID when the ID file is full.
CORRECTIVE ACTION: Determine the reason for the error display and take the appropriate
corrective action.
HUH?
UNITS
key pressed in an attempt to perform a “unit” conversion that is not allowed.
CORRECTIVE ACTION: Determine the reason for the error display and take the appropriate
corrective action.
ID IN USE
The operator is attempting to add an ID that already exists.
CORRECTIVE ACTION: Chose another number for the ID.
ID NOT FOUND
The operator is attempting to use an ID that does not exists.
CORRECTIVE ACTION: Determine the reason for the error display and take the appropriate
corrective action.