15
SETUP AND CALIBRATION
Your 220 indicator has been thoroughly tested and
calibrated before being shipped to you. If you
received the indicator attached to a scale,
calibration is not necessary. If the indicator is
being connected to a scale for the first time or
recalibration is necessary for other reasons,
proceed as indicated.
The calibration switch is located on the enclosure
rear panel. You may gain access to this switch
simply by removing the calibration switch access
screw on the rear panel. Refer to Figure No. 8.
Figure No. 8
During the setup and calibration process it is necessary to enter operational parameters via the
220's keypad. Pressing the
ENTER
key without
entering a new value will retain the current
setting and advance the 220 to the next prompt. To change a setting, enter a new value and
press the
ENTER
key. This will save the new value and advance the 220 to the next prompt.
Pressing the
ASTERISK
key will "backup" to the previous prompt or return to normal operation.
CAUTION: The membrane keypad is not to be operated with pointed objects
(pencils, pens, fingernails, etc). Damage to keypad resulting from this practice will
NOT be covered under warranty.
Enter Setup Mode
To enter the setup mode, with the indicator ON, insert a small screwdriver or other tool through the
calibration switch access hole on the rear panel. Press and release the calibration switch. The
menu
SETUP
will be displayed. Continue to press and release (or hold) the switch to rotate through
the beginning point for entering the setup mode.
SETUP Setup Mode (starts at USA= prompt)
ANALOG
Analog to Digital Filtering (starts at
FILTER=
prompt)
CALIBRATE
Calibration (starts at
CAL WT 1=
prompt)
SERIAL Serial Input/Output (starts at BAUD 1= prompt)
PRINT TABS
Print Tab Settings (starts at
PORT=
prompt)
FINE SPAN Fine Span Adjustment
HIGH RES
Display High Resolution Weight Mode
KEY LOCK OUT
Key Lock Out Function
ANALOG OUT Analog Output Option (If option board is installed, Calibration
of 10 volt or 4 to 20 mA Analog Output)
If you press the ENTER key at the
SETUP
prompt, you may proceed through to the next section (up
to and including
FINE SPAN
) by pressing the ENTER key.
NOTE! Setup may be interrupted at any time. ALL data previously entered and
finalized with the ENTER key will be retained in the non-volatile memory.
Pressing the calibration switch
at any prompt
will return you to the
SETUP
menu. To exit setup,
press the ASTERISK key with any of the above menu selections displayed or cycle power at any
time (press the ON/OFF key twice).
Calibration Switch
Access Screw