2011-03-31 9G3886_en 5-1
5
Specifications
Equipment specifications
Operating space requirements
Allow 5 cm (2 in.) clearance around the top, sides, and back of the laser
imager. This space is required to let the laser imager perform normal
operator functions. For service functions, allow 15 cm (6 in.) clearance
around the laser imager.
Environmental requirements
Temperature Operating: 15° to 33° C (59° to 91° F)
Storage: -40° to 55° C (-40° to 131° F)
Relative humidity Operating: 20% to 85% RH, noncondensing
Storage: 10% to 90% RH, noncondensing
Altitude 30 m (100 ft) below sea level to 3,048 m (10,000 ft) above sea level
Surface levelness The countertop/floor must be level within 1°.
Environmental effects
Acoustical Noise:
Less than 55 dB at 1.5 m during idle or standby
Less than 75 dB momentary at 1 m during normal operation
Unpacked Packed
Height 47 cm (19 in.) 74 cm (29 in.)
Width 61 cm (24 in.) 79 cm (31 in.)
Depth 66 cm (26 in.) 85 cm (34 in.)
Weight 54 kg (120 lbs) 73 kg (160 lbs)