29
Menu-GEN.09: Set Zero Key Range
Function Zero Key Range Setting
Display Part Meaning
Set Range
(0~99)
ZKR. XX
Initial Value :
10 %
Zero operates up to within +/- ᄋᄋ% of the maximum weight
Menu-GEN.10: Set Tare Key Range
Function Tare Key Range Setting
Display Part Meaning
Set Range
(0~100)
TKR. XX
Initial Value :
100 %
Tare operates up to within +/-ᄋᄋ% of the maximum weight
Menu-GEN.11: Set Initial Zero Range
Function Initial Zero Range Setting
Display Part Meaning
Set Range
(0~99)
ZIR. XX
Initial Value :
10%
Initial zero operates within +/-ᄋᄋ% of the Gross Weight
Menu-GEN.12: Set Overload Range
Function Overload Range Setting
Display Part Meaning
Set Range
(0~99)
OVRD. XX
Initial Value :
9 x 1 Digit
Overweight from the next to 0 x 1 Digit of the maximum weight
Menu-GEN.13: Set Hold Type
Function Hold Type Setting
Display Part Meaning
HOLD.00
Average Value Hold
HOLD.01 Peak Hold
HOLD.02 Sampling Hold
Set Range
(0 ~3)
HOLD.03 Automatic Hold