12-2. Error Message from the Weighing Mode
Error Cause Solution
Err 01
The initialization of the scale cannot be done
because of the shaking weight.
Turn on the power after placing the scale at a flat
p
lace with no vibration.
Err 02
Either the connection of load cell is wrong or
there is abnormality in the A/D conversion
section.
Check the connection between the load tray and the
body.
Err 05
Either you are pressing a key for a long time or
there is abnormality in the key section.
Make an inquiry to A/S.
Err 08
The zero key, tare key and start key were
disabled at the instable weight.
Set the zero key, tare key and start key to the proper
user conditions at F14 in the Set Mode.
Err 09
The current weight is out of the range of zero
p
oint.
Set the range of operations for the zero key to within
2% or 10% at F13 in the Set Mode.
Err 10
The tare to set is out of the maximum weight of
the scale.
Set the tare to less than the maximum weight.
Err 12
The type of the configured printer is one that
cannot support the total print.
DLP printers cannot make the total print.
Set "F40" to '2' when DEP printers are used.
Err 13
The set value of zero point on the weight setting
is out of range.
Check the status of the load tray and set the weight
again.
Err 15
The range exceeded during setting the item code
in the command mode.
Check the range of item code.
Err 82
There is abnormalities in the A/D set section.. Make an inquiry to A/S.
Over
The current weight on the load tray is too heavy
and out of the allowable tolerance.
Avoid any weight in excess of the maximum
allowable limit on the scale.
If the load cell is damaged, it should be replaced.