CAL 5
FUNCTION: Span Calibration
KEY
DISPLAY DESCRIPTION
ENTER(LIGHT) :
Span calibration and
Move into next menu.
CAL 5 condition
Load the weight which was set in
CAL 3 and press LIGHT.
Under span calibration.
Span calibration is completed.
Press ENTER(LIGHT) key.
(Save & Exit CAL Mode)
REF 1. If Span calibration is done without any error, GOOD message is displayed
The weight of setting weight is displayed on LCD screen.
Check the weight.
REF 2. If the span is low, Error message (ERR 24) is displayed.
Calibrate with lower resolution.
4. SEALING METHOD
Install the seal on the wire loop as shown in below Figure.
4