36
(2) ERRORS IN CALIBRATION MODE
Err 21
▣ Reason
The resolution is set to be exceeded the limit 1/10,000.
☞ Trouble shooting
Lower the resolution.
The resolution = allowed weight/one division
Modify the allowed weight in CAL1 or modify the division in CAL2 so that the
resolution should be below 1/10,000.
Err 22
▣ Reason
The weight for span calibration is set to be lower than 10 % of the maximum
capacity of the scale.
☞ Trouble shooting
Set the weight for span calibration in CAL3 to be more than 10% of the maximum
capacity.
Err 23
▣ Reason
The weight for span calibration is set to be exceeded 100 % of the maximum
capacity of the scale.
☞ Trouble shooting
Set the weight for span calibration to be within the maximum capacity of the scale
in CAL 1.