35
Err 10
Reason
Tare weight exceeds the maximum capacity of the scale.
Trouble shooting
Set the tare to be smaller than the maximum capacity.
Otherwise the maximum capacity is reset to be larger than the tare to be set in the
calibration menu, and reset the calibration using weight.
Err 13
Reason
The zero range deviates from the set range.
Trouble shooting
Confirm that there is nothing on the weighing platform.
If nothing exist, do calibration in CAL mode.
Over
Reason
The weight on platform is too heavy to be measured.
Trouble shooting
Do not load the item exceeds the maximum tolerance.
If the load cell is damaged, the load cell should be replace.