CAS CI-2001B - Err 22: Low Span Calibration Weight; Err 23: High Span Calibration Weight; Err 24: Small Load Cell Output (Span); Err 25: Large Load Cell Output (Span)

CAS CI-2001B
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
Err 24
Reason
The load cell output is too small at SPAN calibration.
Trouble shooting
Setting of current resolution is not possible due to the error in load cell. Proceed
calibration again with less resolution.
Loadcell Sense Voltage
for 5V Excitation Voltage
Recommended Resolution
2 mV 1/1,000
4 mV 1/2,000
10 mV 1/5,000
Err 25
Reason
The load cell output is too large at SPAN calibration.
Trouble shooting
Setting of current resolution is not possible due to the error in load cell. Proceed
calibration again with less resolution.
Err 26
Reason
The load cell output is too large at ZERO calibration.
Trouble shooting
Check whether the platform empty.
Proceed calibration again after checking in A/D TEST mode.