Menu-2106
Function Set Stable Range
Display Part Meaning
Set Range
(0 ~99)
ᄋ x 0.5 division
Initial Value:
1x 0.5 division
Stability lamp is turned ON when weight change is such that
the width of change in a given time is within the set value x 0.5
division
Note 1. Function that acknowledges it as the stable state when the width of weight change
within a set time does not exceed the set range X 0.5 division.
Note 2. Weighing stabilization will be made faster by setting the larger number if the
environment involves much vibration in the surrounding and by setting the smaller
number if there is little vibration.
Menu-2107
Function Set Automatic Zero Tracking Compensation
Display Part Meaning
Set Range
(0 ~ 99)
ᄋ x 0.5 division
Initial Value:
1x 0.5 division
Function to compensate for zero when weight change is
such that the width of change in a given time is within the
set value x 0.5 division
Note 1. This function automatically calibrates for zero if the weight does not exceed a given
range of division within a given time in the zero state.
Ex) When the maximum display division is 120.0Kg with the value of one
division set as 0.05Kg, provided that the Menu[2-1-07] is set as “2”,
If there is a weight change by less than 100g
within 0.5SEC, the indicator automatically
dis
la
s it as “0”.
If there is
a change by more th
100g within 0.5SEC. a chang
0.5 Sec
2
.0digit
(0.1Kg)
an
ed
5