Menu-2111
Function Set Hold Canceling Conditions
Display Part Meaning
ᆷ 1_Cancel Hold
at zero
Hold is canceled when it becomes zero.
ᆷ 2_Cancel upon
entering Hold Key
Hold is canceled when Hold key is entered.
Set Range
(1~3)
ᆷ 3_Cancel upon
entering Hold less
Key
Hold is canceled when Hold less key is entered.
Menu-2112
Function Set Automatic Hold Starting Conditions
Display Part Meaning
Set Range
(0, 99)
ᄋ x 1 division
Initial Value:
10x 1 divistion
Hold starts when the weight changes within the set range
value x 1 division.
Menu-2113
Function Set Automatic Hold Canceling Conditions
Display Part Meaning
Set Range
(0~99)
ᄋᄋ%
Initial Value: 10 %
Hold is canceled when the value is changed by more
than ᄋᄋ% of the held value.
Menu-2114
Function Set Ker Operating Conditions (ZERO, TARE Keys Availability)
Display Part Meaning
ᆷ 1 _ Always
Operational
Always in operation
Set Range
(1, 2)
ᆷ 2_ Operational
when the weight is
stable
Operates only if the weight is stable
Menu-2115
Function Set Zero Key Range
Display Part Meaning
Set Range
(0~99)
ᄋᄋ %
Initial Value: 1 %
Zero key operates up to within +/- ᄋᄋ% of the maximum
weight
56