Appendix 2> Command Mode 2 Description
CAS <NT
-570 Command>
Command data to NT-570A
0 1 2 0 1 2 0 1 2 0
1 2
Command description NT-570A Respond
D ID K Z CR LF ZERO key
D ID K T CR LF TARE key Return the received
D ID K G CR LF GROSS key Return the received
D ID K N CR LF NET key Return the received
D ID K S CR LF START key Return the received
D ID K P CR LF STOP key Return the received
D ID K B CR LF Print key Return the received
D ID K C CR LF Total print key Return the received
D ID K W CR LF Request weight data Return the received
D ID H T CR LF Request set point value Send Format 2
D ID S 1 0 0 0 0 0 CR LF 1
st
Step value Return the received
D ID S 2 0 0 0 0 0 CR LF 2nd Step value Return the received
D ID S 3 0 0 0 0 0 CR LF 3rd Step value Return the received
D ID S 4 0 0 0 0 0 CR LF 4th Step value Return the received
D ID S 5 0 0 0 0 0 CR LF High limit value Return the received
D ID S 6 0 0 0 0 0 CR LF Low limit value Return the received
D ID H E 0 0 0 0 0 CR LF Set point code(00-99) Return the received
(D, ID:00~99, CR : 0×13, LF: 0×10, Command HC, HE range = 00 ~ 99)
* Format 1 : PC send set point all data to indicator NT-580A
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
D ID H A Set point code , Zero Band , Optional-
20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39
Preli , Preliminary , Final value , Free Fall
40 41 42 43 44 45 46 47 48 49 50 51 52 53
, High limit , Low limit C
LF
* Format 2 : Recieve the request data from PC then response of Indicator
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
D ID H T Set point code , Zero Band , Optional-
20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39
Preli , Preliminary , Final value , Free Fall
40 41 42 43 44 45 46 47 48 49 50 51 52 53
,
High limit
,
Low limit
CR LF
* Please input without the decimal point.
7