EasyManua.ls Logo

CAS DH - Fast Calibration Methods

CAS DH
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
(2)Press [HOLD] key during startup initialization, the indicator will come into calibration
state.
Fast Zero calibration:
Press [FUNC] key in any time before display nOLOAD,the indicator will save the
parameter of division, decimal point, full capacityand come into the zero calibration state
directly. Press [ZERO] key when the stabilization symbol appear, it displays End】,it
means that the indicator save the full capacity calibration parameter before and press [CLEAR]
key, the indicator save the parameters and back to weighing state.
Come into full capacity calibration state directly
Press [ACCUM] key in any time before display AdLOAd,the indicator will save the
parameters of division, decimal point, full capacityand come into the full capacity calibration
state directly.
Note Pull out short-circuit ring from JP7 after the calibration, and lead sealing again.

Related product manuals