13
(2)Press [HOLD] key during startup initialization, the indicator will come into calibration
state.
Fast Zero calibration:
Press [FUNC] key in any time before display 【nOLOAD】,the indicator will save the
parameter of division, decimal point, full capacity,and come into the zero calibration state
directly. Press [ZERO] key when the stabilization symbol ▼ appear, it displays【 End】,it
means that the indicator save the full capacity calibration parameter before and press [CLEAR]
key, the indicator save the parameters and back to weighing state.
Come into full capacity calibration state directly:
Press [ACCUM] key in any time before display 【AdLOAd】,the indicator will save the
parameters of division, decimal point, full capacity,and come into the full capacity calibration
state directly.
Note :Pull out short-circuit ring from JP7 after the calibration, and lead sealing again.