16 Troubleshooting
59
16.2 Additional Errors
A/D Converter Error-Check connection between load cell and A/D converter
Replace load cell
Replace A/D converter
NV Fragmentation Error-Perform NV defragmentation in Calibration menu
Calibration Warning-Recalibrate Scale-NOTE: often when this message appears
You cannot get into Calibration Mode by holding ON/OFF while powering on. In this
case, you must use the calibration switch on top of the main cpu.