2
4
CAL 8
Function : Zero Adjust
KEY VFD SCREEN DESCRIPTION
2-CAL
Unload the tray and press the Enter Key
1234
Current A/D Value displayed
Check the stable and press the Enter Key.
”* " : Cancel and
move to CAL-1
Enter : Store and move
into next menu
- - -
Under Zero Calibration
Ref 1. Use it that the initial zero range can not be passed because the load cell is damaged
Ref 2. If the zero Calibration is done without any error, the program moves into CAL-1 automatically.
Ref 3. If the zero value is too low, the error message “Err 26” is displayed.
Ref 4. If the zero value is too high, the error message “Err 27” is displayed.
CAL 9
Function : Weight Constant Calibration
KEY VFD SCREEN DESCRIPTION
NOT USE
You can use this function because of the Multi-
Calibration.
FACtor
It means the weight constant
calibration Mode
▲,▼ : increase or
decrease of no.
, : Shift of digit▶◀
”* " : Cancel and
move to CAL-1
Enter : Store and move
into next menu
12345
Current Factor Value displayed
Ref 1. The ordinary user doesn’t need to use this menu because it is used for the Calibration without a
weight.
Ref 2. It is availble only when setting the range of the multi-Calibration to 1.
When setting the range of CAL4-1 over 2, “NOT USE” is displayed.
Ref 3. If you want to enter the weight constant calibration Mode , you must input the password then.