41Method - Create, Edit and Delete
proportion of each product if there are 2 or more products being blended. The sum of all the propor-
tions must equal 100.00% (50%/50%, 75%/20%/5%, etc).
6.5. If the product has any additional additives (such as added water, salt, spices, etc), these should be
added in the Bias section. Similarly, if after running 3-5 samples there is a consistent bias present
between the ProFat results and the reference results, a bias can be added for that purpose.
CAUTION
Not following proper sample preparation and spread techniques can lead to a reduction in precision,
inaccurate results, or even damage to the system.
WARNING
Improper sample preparation and spread techniques can create an unsafe testing environment due to re
or volatiles in the cavity.