MAINTENANCE
EN
|
Intelect
®
Mobile 2 ELECTROTHERAPY User Manual
CLEANING THE INTELECT
®
MOBILE 
With the system disconnected from the power source, clean
the system with a clean, lint-free cloth moistened with
water and mild antibacterial soap. If a more sterile cleaning
is needed, use a cloth moistened with an antimicrobial
cleaner. Cleaning should be performed daily.
Do not submerge the system in liquids. Should the unit
accidentally become submerged, contact the dealer or DJO
Service Department immediately.
Cleaning the LCD Screen
Clean the LCD with a clean, dry cloth, in the same way as
cleaning the computer monitor screen. Do not use abrasive
materials or chemicals or liquids.
VACUUM MODULE CLEANING
Reservoir Draining
When draining the reservoir, wear surgical type
gloves. To drain the reservoir cup, rotate clockwise
direction, shown below. Dispose of contents in
accordance with your national, state or local disposal
guidelines.
Flushing Lead Hoses and Reservoir
1. Connect all two lead hoses to the Vacuum Module.
Submerse the other end of the lead hoses into a
container filled with at least  ml ( fl oz) of hot water
and one drop of dishwashing detergent added.
2. Turn Vacuum Module ON and set Vacuum Intensity to
maximum.
3. Repeat this procedure until no particles are visible when
the reservoir is drained.
4. Dispose of reservoir contents according to national,
state and local rules and regulations.
5. The Vacuum system should be flushed weekly.
Note: The vacuun system should be flushed and drained prior to
vacuum module storage or transport if temperature could potentially
be below 0°C
Cleaning the Vacuum Module
With the system disconnected from the power
source, clean the vacuum module with a clean,
lint free cloth moistened with water and mild
antibacterial soap. If a more sterile cleaning is needed,
use a cloth moistened with an antimicrobial cleaner.
Do not submerse the vacuum module in liquids.
Should the vacuum module accidentally become
submersed, contact the dealer or DJO Technical
Service Department immediately.
Cleaning Instructions for the Electrodes and Suction
Cups
A mild antibacterial solution containing no chlorine
can be administered with a cloth and wiped or
air dried. This is recommended between patient
treatments. These electrodes are multiuse when
properly maintained and cleaned.
Cleaning Instructions for the Sponges
The associated sponges are recommended for
single patient use only and should be cleaned with a
% alcohol solution before and after each therapy
session.

Table of Contents

Other manuals for Chattanooga Intelect Mobile 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chattanooga Intelect Mobile 2 and is the answer not in the manual?

Chattanooga Intelect Mobile 2 Specifications

General IconGeneral
BrandChattanooga
ModelIntelect Mobile 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals