SYSTEM
EN
|
Intelect
®
Mobile 2 ELECTROTHERAPY User Manual
POWER
Input 100 - 240 V AC, 1.0 to 0.42 A, 50/60 Hz
Electrical Class CLASS II
Mode of Operation Continuous
Note: Mains isolation is achieved by use of the double pole switch
located on the rear panel.
Electrical Type (Degree of Protection)
Electrotherapy .TYPE BF
Electrotherapy Vacuum .TYPE BF
ELECTRO STIMULATION SPECIFICATIONS
Output specifications are described for each waveform from
pages 24-26.
Unless otherwise specified, electrotherapy controls accuracy
is: ± 20 %.
Load impedance: 500-1000 Ohm
CC = constant current, eect of load impedance on voltage
CV = constant voltage, eect of load impedance on current
SYSTEM SPECIFICATIONS AND DIMENSIONS
Width Depth Height Weight (no battery)
Intelect Mobile 2 Head Unit
STIM .cm .cm cm .kg
Cart configurations
Cart (Safe working load 6.5kg) cm MAX cm MAX cm .kg
Cart with vacuum cm MAX cm MAX cm
.kg
Device on cart - - cm -
VACUUM SPECIFICATIONS
Power
Input 20-25 Vdc, maximum peak current 4A
Electrical Type TYPE BF
General characteristics
Vacuum Range .0 to 600 mbar maximum +/- 5%
Vacuum Modes Continuous or Pulsed Continuous.
10 setting over vacuum range,
60 mbar per setting,
+10 mbar to 10 mbar per setting
Pulsed Mode
Maximum Vacuum settings 2 to 10, +10mbar to -10mbar
per setting
Minimum Vacuum settings in 1 to 9, +10mbar to
-10mbar per setting
Hold Time in minimum & maximum vacuum settings,
0-20 seconds, in 1 second steps, +/-0.5 seconds